CAS would like to make our customers aware that Vaisala is issuing a product recall for the old models of the Veriteq compact lithium battery powered data loggers. Veriteq was purchased by Vaisala in 2010. The loggers in question were manufactured by Veriteq Instruments and sold as either the Veriteq Spectrum/Model 5 or private labeled by dataTaker as the DT1000-4000 Series.
In very rare cases, these loggers are subject to potentially experiencing overheating when using the device connected to a network via an external serial or USB cable. The serial line connection may unintentionally start to recharge the lithium-thionyl-chloride battery inside the Data Logger. Although very unlikely, the overheating may in some instances cause the battery to breach its casing.
The affected product models were sold by Veriteq as the VL1000, SP1000, VL14, SP1400, VL1700, SP1700, VL2000, SP2000, VL4000 and SP4000 and by dataTaker as the DT1000, DT1700, DT2000, and DT4000. The products can be identified by the physical appearance and a Veriteq or Data Electronics/Datataker label. They can be identified also by the serial numbers 00 000 001 through 13 999 999 and 97 000 001 through 99 999 999. The affected model Data Loggers have rectangular, 10-pin Harwin serial ports for connection instead of a standard D-connector. The recalled devices were produced during 1997-2013.
Required Actions
To prevent the risk of the battery overheating, disconnect any Data Logger that is connected to any external network or serial converter. When removing the Data Logger from your location, if visual inspection shows damage or leaking residue, please wear protective garments—i.e. safety glasses and gloves.
Do not ship the units back to CAS DataLoggers or Vaisala. Please discard them as per the battery manufacturer Tadiran’s Safety Notice here.
Corrective Action
Customers who have Veriteq Spectrum or equivalent dataTaker DT1000-4000 Series Data Loggers in use can have these units replaced with a current production version.
Please contact Vaisala Regional Service Centers for support and replacements and have available: the model type and serial numbers of the Data Loggers you will exchange. The new units will have a 2-year warranty and will be delivered with calibration certificates.
Vaisala Regional Service Centers
North America:
Tel: +1 888 824 7252
Email: helpdesk@vaisala.com
Japan:
Tel: +81 3 5259 5960
Email: japan.support@vaisala.com
Other Countries:
Email: helpdesk@vaisala.com
###
Source: https://www.dataloggerinc.com/news-article/recall-notice-for-veriteq-data-loggers/